Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Astro-Med PM02: (1 year) CAL VER /5520 DATE: 26-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 11 NUMBER OF LINES: 92 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5520/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: 20 min. - REQUIRED 1.005 DISP Temperature: 18 - 28 degC 1.005 DISP Humidity: <75% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD INITIAL SETUP 1.008 DISP Press the ZERO DEFAULT. 1.008 DISP 1.008 DISP Ground Channels. 1.008 DISP 1.008 DISP Set for Bottom Baseline. 1.008 DISP 1.008 DISP Select Peak to Peak Mode. 1.009 HEAD {FULL SCALE AMPLIFIER} 1.010 DISP Set to 100mV range. 1.011 HEAD FULL SCALE AMPLIFIER: {100mV Range} 1.012 5520 100mV S 2W 1.013 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 2.001 HEAD FULL SCALE AMPLIFIER: {250mV Range} 2.002 DISP Set to 250mV range. 2.003 5520 250mV S 2W 2.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 3.001 HEAD FULL SCALE AMPLIFIER: {500mV Range} 3.002 DISP Set to 500mV range. 3.003 5520 500mV S 2W 3.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 4.001 HEAD FULL SCALE AMPLIFIER: {1V Range} 4.002 DISP Set to 1V range. 4.003 5520 1V S 2W 4.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 5.001 HEAD FULL SCALE AMPLIFIER: {5V Range} 5.002 DISP Set to 5V range. 5.003 5520 5V S 2W 5.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 6.001 HEAD FULL SCALE AMPLIFIER: {10V Range} 6.002 DISP Set to 10V range. 6.003 5520 10V S 2W 6.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 7.001 HEAD FULL SCALE AMPLIFIER: {50V Range} 7.002 DISP Set to 50V range. 7.003 5520 50V S 2W 7.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 8.001 HEAD FULL SCALE AMPLIFIER: {100V Range} 8.002 DISP Set to 100V range. 8.003 5520 100V S 2W 8.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 9.001 HEAD FULL SCALE AMPLIFIER: {250V Range} 9.002 DISP Set to 250V range. 9.003 5520 250V S 2W 9.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 10.001 HEAD {FREQUENCY RESPONSE} 10.002 DISP Set the trace to center scale. 10.002 DISP 10.002 DISP Set to 10V range. 10.002 DISP 10.002 DISP Set to 50 division grid. 10.003 5520 10V 25kH SI S 2W 10.004 EVAL Is {display >35 division (-3dB point)}? 11.001 END